|
Carbonates
Carbonates are analysed using a Micromass Multiprep device. Samples are loaded into 1mL Weaton V-vials which are sealed with Kel-F and silicone septa and placed in a heated block maintained at 90oC. A double-bore needle is pushed through the septa and into the vial, the vial is evacuated and then pure phosphoric acid (103%) is dispensed onto the sample. The evolved CO2 is passed through the water trap and cryogenically collected in the external cold finger. After all the gas has been collected it is released into the sample side of the dual inlet and is analysed against the working gas. Calibration is performed against an internal laboratory standard, NBS18 and NBS19. The routine minamum sample size is about 30mg of calcium carbonate, although it is possible to analyse samples as small as 20mg.
Oxygen in Water
Sample aliquots (200mL)are loaded into 1mL Weaton V-vials. Using a glove bag, the vials are filled with CO2 and then sealed with Kel-F and silicone septa and placed in a heated block maintained at 40oC. The samples are equilibrated for 5h. prior to analysis to ensure that the d18O of the CO2is identical to that of the water. During analysis the needle is pushed into the silicone septum and the transfer capillary is evacuated. Then the needle is pushed through both septa to sample the CO2 in the headspace above the water. The CO2is passed through the water trap and into the sample side of the dual inlet and is analysed against the working gas. Calibration is performed against internal laboratory standards and V-SMOW, GISP and SLAP (as appropriate).
Hydrogen in Water
Sample aliquots (200mL)are loaded into 1mL Weaton V-vials with a few 'hokko beads'. The vials are sealed with Kel-F and silicone septa and placed in a heated block maintained at 40oC. The needle is pushed through the septa, the vial is evacuated and then filled with hydrogen, and the needle removed. After equilibrating for 2h. to ensure that the dD of the H2 is identical to that of the water the needle is re-inserted to sample the hydrogen in the headspace above the water. The hydrogen is passed through the water trap and into the sample side of the dual inlet and is analysed against the working gas. Calibration is performed against internal laboratory standards and V-SMOW, GISP and SLAP (as appropriate).
Carbon or Nitrogen in Combustible organic samples
Aliquots of powdered sample are weighed into tin cups and the cups are crushed to exclude air. The autoloader drops the cups into the combustion tube which is heated to 1050oC while oxygen is injected into the helium carrier gas. The tin and sample flash combust and the product gasses are passed over catalysts and high purity copper. This ensures that all carbon is present as CO2 and all nitrogen as N2. Water is removed by a magnesium perchlorate trap. For nitrogen analyses the CO2 is removed by a 'Carbasorb' trap.
The product gas is carried to the mass spectrometer for measurment of the 13C/12C ratio relative to a reference gas. Calibration is achieved by including standard reference materials in the run at regular intervals
|